Connect to an Intubated Patient

1.

Connect the gas sampling elbow and elbow
connector into the patient’s breathing circuit
according to the manufacturer’s instructions.

2.

Connect the sample line to the elbow connector and the watertrap. Make sure the
sample line is firmly connected.

CO

2

Messages

Mainstream CO

2

Messages

Equipment messages for the Mainstream CO

2

option can appear on the display in an

equipment alert window and in numeric zones.

If a sensor is damaged, contact Welch Allyn’ Technical Services Department for information
on sensor service options.

ALTIMETER FAILURE - RANGE

. The Propaq is operating at an altitude outside the

Mainstream CO

2

option’s operating altitude range of -2,000 to 15,000 feet. Returning the

monitor to within this range automatically cancels this message and restores operation.

Warning The exhaust port for Sidestream

CO

2

is an output for the expired gases from

the patient and any connected breathing
apparatus. The exhaust port is intended only
for connection to gas collection equipment
such as gas scavenger devices. Do not
allow any other connection to the
exhaust port.

If you use a gas scavenging system with
Sidestream

CO

2

, be sure to install it

according to the manufacturer’s
instructions. The scavenging system
should comply with ISO 8835-3:1997 (E).

Sidestream

CO

2

accuracy decreases if additional tubing is connected to the

sample line. Avoid connecting additional tubing to the standard sample line.

Exhaust Port.

WARNING - Do not connect sample

line or patient input to exhaust port.

Watertrap

Sample Line

Elbow

Connector