Temperature

Propaq Encore Reference Guide

143

Welch Allyn

Temperature

Temperature Specifications

Minimum Time between automatic
measurements

30 seconds (Auto Mode)
2 seconds (Turbo Mode)

Electrosurgery Interference
Suppression

Included in all models: 202EL, 204EL, 206EL

NIBP Performance

Per EN 1060-1, EN 1060-3, and ANSI-AAMI SP10-1992

NIBP Safety

Per IEC 601-2-30

Characteristic

Specification

Range

0° to +50°C; 32° to +122°F

Displays

T1, T2, and

T

Probes

Compatible with YSI Series 400 and 700 probes. HP side panel only compatible with
YSI 400 and has HP connector.

Units

°C and °F selectable

Channel Accuracy

Temperature Range

Tolerance

0° to +10°C
>10° to +50°C

+32° to +50°F
>50° to +122°F

±0.2°C
±0.1°C

±0.4°F
±0.2°F

Resolution

0.1°C or °F

Electrosurgery
interference
suppression

Included in all models: 202EL, 204EL, 206EL

Characteristic