Responding to patient alarms, Suspend an alarm tone, Turn off or change alarm limits – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 95

background image

Types of Propaq Alarms

Propaq Encore Reference Guide

95

Welch Allyn

Responding to Patient Alarms

Suspend an Alarm Tone

You can temporarily turn off an alarm tone by pressing the

SUSPEND/RESUME

key to the left of

the screen or the

SUSPEND

button in the Patient Alarm Menu. The tone is suspended for 90

seconds. During that period, the visual alarm indications continue. When you suspend the
alarm, the red ALARM light is on.

You can “unsuspend” the alarm before 90 seconds has elapsed by pressing the

SUSPEND/

RESUME

key to the left of the screen, or the

RESUME

button in the Alarms Menu. If an alarm

condition still exists, the tone will again sound. For NIBP (both pressure values and pulse
rates derived from NIBP), pressing

RESUME

will not resume the NIBP alarm because NIBP is

not continuously measured.

If the

LIMITS

button is available, you can make Propaq Encore alarm limit adjustments by

pressing

LIMITS

.

Turn Off or Change Alarm Limits

Possible adjustments in order to cancel the alarm are:

Turn off the alarm limits only for the violating alarm limit values by pressing

ON/OFF

in the Limits Menu after you have selected the appropriate vital sign limits window
(

LIMITS

,

NEXT PAGE

,

ON/OFF

)

Warning Suspending an alarm suspends all alarm tones for 90 seconds or
until the

RESUME

button is pressed.

Flashing numerics
indicate the limits that
have been violated.

Illuminated

alarm light

Patient Alarm

Menu