Types of Propaq Alarms

Propaq Encore Reference Guide

93

Welch Allyn

Types of Propaq Alarms

Patient Alarms: Definitions and Indications

The Propaq Encore can alert you to changing patient conditions through its programmable
alarm functions. After you set alarm limits, any vital sign that violates any of its limits results
in both audible and visual alarm indications.

The Propaq Encore also shows you when any alarm limit is turned off by illuminating the
amber

ALARM(S) OFF

light.

A steady, high-pitched alarm tone sounds whenever a patient alarm limit is violated.

The alarm tone continues until:

the patient condition changes and no longer violates the limit

you suspend the alarm tone by pressing the

SUSPEND

button

you adjust the alarm limit so the vital sign does not violate it

you turn off the violated alarm limit

Patient and Alarm Limit Status

Red

ALARM Light

Yellow

ALARM(S) OFF

Light

Tone

Patient in alarm condition, and all alarm
limits are on

FLASH

OFF

ON

Patient in alarm condition, and at least one
alarm limit is off

FLASH

ON

ON

Patient alarms suspended (whether in
alarm condition or not)

ON

ON or OFF

OFF

Patient not in alarm condition, and at least
one alarm limit is off.

OFF

ON

OFF

Equipment alert, patient not in alarm
condition

OFF

FLASH

ON

for 1 second,

OFF

for 4 seconds

Equipment alert, patient alarms suspended

ON

FLASH

OFF