Factory Default Settings
Propaq Encore Reference Guide
159
Welch Allyn
Display Wave Select
All waves are on except NIBP
Trend Group
NIBP
Alarm Limits
All are ON except P2
HR Limits
Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic
Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean
Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
P1, P2 Limits - Diastolic
Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
P1, P2 Limits - Mean
Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
SpO
2
Limits
Adult: 85%, 100%
Ped: 85%, 100%
Neonate: 80%, 95%
RR/BR
Adult: 5, 30 BrM
Ped: 10, 45 BrM
Neonate: 10, 60 BrM
TEMP Limits
35.0°, 37.8° C
∆T Limits
0.0°, 2.8° C
ETCO
2
Limits
25, 60 mmHg (3, 8 for % and kPa)
INCO
2
Limits
N/A, 5 mmHg (0.7 for % and kPa)
Apnea Delay
Adult/Ped: 20 seconds
Neonate: 15 seconds
PRINTER SETTINGS
Printer Alarm Print
OFF
Printer Auto Print
OFF
Printer NIBP Ticket
OFF
Printer Apnea Ticket
ON
Printer Print Speed
25 mm/sec
Printer Auto Trend
OFF
Printer Trend Selections
NIBP and P1 = ON; all others = OFF
Printer OxyCRG on Alarm
OFF
a.Any time you change the Date, Filter, Temp F/C, Decimal, or CO
2
Units setting, the new setting also becomes
the powerup default setting.
Characteristic
Specification