Bio-Rad Bio-Plex Software® Upgrades and Conversions User Manual
Page 145

Calibration, Validation, and Instrument Operations Logs
133
When you click OK, a Save As dialog opens.
Figure 117. Save As dialog for Secure Protocol files
Enter a name for the file or use the default name and click Save.
After a file has been signed, it is flagged as read-only and cannot be modified.
Any subsequent changes must be saved as a new file. Also, any changes
made to a file after it has been signed will generate an audit trail documenting
each change, as described starting on page 137.
A new file generated from a signed file can be signed again to verify and lock
in any changes made to it. As before, the new signed file is flagged as read-
only and any subsequent changes must be saved under a new file name.