Metrohm 789 Robotic Sample Processor XL User Manual

Page 93

background image

3.4

Commands

Metrohm Sample Processor, Operation

85

>sample sequence

13 DOSRATE 1

160

1…3

0.01…160 mL/min

Dosing rate

The dosing rate can be set individually for all three dosing devices. The

first parameter stands for the address (MSB1…3) of the dosing device,

the second parameter allows the setting of the dosing rate in mL/min.

>sample sequence

13 FILLRATE 1

160

1…3

0.01…160 mL/min

Filling rate

The filling rate can be set individually for all three dosing devices. The

first parameter stands for the address (MSB1…3)of the dosing device,

the second parameter allows the setting of the filling rate in mL/min.

>sample sequence

13 COCKMOVE 1 : auto

1…3

auto,

not over, desc.,

asc.

Direction of rotation of the stop-

cock of a Dosing unit

The direction of rotation of the stopcock can be defined for each Dos-

ino. Apart from the automatic selection (shortest path) a descending or

ascending direction can also be selected.

Not over

means that rotation

will never be carried out over the protected port (see Section 3.3.4).

>sample sequence

13 LIFTRATE 1

25 mm/s

1,2 5…25 mm/s

Lift rate

The lift speed can be set individually for both towers. The first parame-

ter stands for the selection of the tower, the second parameter allows

the setting of the lift speed in mm/s.

>sample sequence

13 SHIFTRATE: auto. 20

auto.,+,–

3…20 w/s

Direction and speed of rotation

of the rack

Direction of rotation and speed of rotation of the sample rack can be al-

tered as required. The first parameter determines the direction of rota-

tion.

auto.

: The Sample Processor automatically selects the shortest path.

+

:

The rack rotates counterclockwise (to a higher rack position)

:

The rack rotates clockwise (to a lower rack position)

The second parameter determines the speed of rotation in degrees/s.