3 connecting the keypad, 4 mounting a 786 swing head – Metrohm 789 Robotic Sample Processor XL User Manual
Page 35

2.2 Instrument setup
Metrohm Sample Processor, Installation
27
2.2.3 Connecting the keypad
778
789
The keypad is attached to the key-
board socket provided for it on the
rear panel of the instrument. To re-
move the plug press both sides to-
gether.
2.2.4 Mounting a 786 Swing Head
The procedure for the installation of the 786 Swing Head is described
below. This is normally carried out by the service technician.
For Sample Processors with 2 towers Tower 1 is moved to a central lift
position and Tower 2 to the parking position.
Switch off the instrument at the mains switch.
• Loosen the titration head holder fitted as standard from the lift
guide by removing the four side-mounted screws.
• Loosen the attachment plate from the titration
head holder and remove the holder.
• Screw the Swing Head onto the attachment
plate with the 2 delivered mounting screws.
• Mount the Swing Head on the lift guide.
Fig. 11 Mounting the Swing Head
Attachment plate
Titration head holder
Lift guide
786 Swing Head