Metrohm 789 Robotic Sample Processor XL User Manual
Page 90

3.4 Commands
82
Metrohm Sample Processor, Operation
CTL
2
CTRL
>sample sequence
9 CTL:Rm START Device1
Rm,RS START Device1
START
Device2
START
Device*
SAMPLE ready
START
Dos1
START
Dos2
START
Dos*
METER Mode pH
METER Mode T
METER Mode U
METER Mode I
METER Mode C
METER Cal pH
METER Cal C
METER
enter
INIT
14 bit (1,0 or ∗)
Setting the Remote lines
= Starts device 1
= Starts device 2
= Starts devices 1+2
= Continue impulse to instrument
= Starts Dosimat at device 1
= Starts Dosimat at device 2
= Starts Dosimat at devices 1+2
= Switches pH meter to pH measurement
= Switches pH meter to temp measurement
= Switches pH meter to mV measurement
= Switches pH meter to Ipol mode
= Switches ion meter to Conc measurement
= Switches pH meter to pH calibration
Switches ion meter to Conc calibration
=
= Initializes Remote interface
any binary pattern with 14 bit
The CTL:Rm command is used for controlling external devices via the
Remote interface. It sets defined line conditions or transmits impulses
via the 14 Remote output lines.
Predefined binary patterns are available; these can be selected via
simple names (e.g.
START device 1
or
METER Mode pH
).
START device x
starts the set mode of a connected Metrohm device.
START Dos x
starts a Dosimat which is connected with a Metrohm titra-
tor via the "activate" line (a special cable is required).
METER xxx
switches to a particular measuring mode with 691, 713, 780 pH Meters
and 692, 781 Ion Meters.
Sample ready
can be used as a continue impulse, e.g. for a connected
Titrando.
Setting special binary patterns allows the flexible monitoring of con-
nected instruments. The following applies
0 = line inactive
1 = line active
∗ = retain line condition
Example:
*************1
= Output line 0 active = start device 1
Details about the Remote interface can be found in the Technical Ref-
erence.