4.3 Error messages

Metrohm Sample Processor, Service, maintenance, errors

121

invalid rack code

The rack code read in by the Sample Processor cannot be

found in the internal table.

missing beaker

After a MOVE command no beaker can be recognized at the

selected position.

rack data missing

No sample rack in position, or no rack data can be found for

the sample rack currently on the turntable.

RS232 error

The transmission parameters of the RS232 interface do not

coincide with those of the receiving device.

SCAN timeout

The connected device has not transmitted the expected sig-

nal within the defined timeout time. The sample determina-

tion may not have been carried out properly, or the connec-

tion has been interrupted. Check the connected device.

* service recommended

The warning limit of the running time meter has been

reached. It is time for the Sample Processor to be serviced.

Please contact your local Metrohm service department.

user memory full

The memory for the user-defined methods is full. Before new

methods can be stored methods that are not used, or only

used infrequently, must be deleted.

wrong rack

The rack placed on the turntable is not the rack assigned to

the method under

parameters

.

trap error xxx

Unexpected program error; switch instrument off and then on

again.

No display, LEDs Tower 1

and Tower 2 light up

LCD error (system error 7). Contact Service.