6 manual stop settings – Metrohm 789 Robotic Sample Processor XL User Manual
Page 80

3.3 Composition of a method
72
Metrohm Sample Processor, Operation
3.3.6 Manual stop settings
The following settings can make a great contribution to the comfortable
operation of the Sample Processor. A definition is made of what is to
happen when the
manual cancellation of a running method, or an emergency stop of the
complete system. Depending on the arrangement of the automated
system and the type of application, it can be laid down in detail how the
individual instrument components and connected instruments are to
react when the
parameters
>manual stop options
Submenu for reaction after manual stop
Open the submenu with
>manual stop
CTL Rmt:
**************
Set signal lines at the Remote interface
Stop device1,
Stop
device2,
Stop
device∗
14 bit (1,0 or ∗)
Connected peripheral devices (e.g. a Metrohm Titrino) can be stopped
automatically. The 14 signal lines of the Remote interface can be set as re-
quired, see also p. 31ff and p. 109ff.
>manual stop
CTL RS232:
Command or character string, for output
via the RS232 interface
14 ASCII characters
default value '&M;$S' (= stops a Titrino)
Connected peripheral devices (e.g. a Metrohm Titrino) can be stopped
automatically. Any character string can be transmitted. For details of the se-
rial RS232 interface see "Technical Reference".
>manual stop
PUMP: OFF
Switches the pumps, or pump connec-
tions
OFF, ON, cont.
(cont. = retain condition)
This setting applies to all pumps and/or pump connections.
>manual stop
STIR T1:
cont.
Switches the stirrer at Tower 1
cont., ON, OFF
(cont.= retain condition)
>manual stop
STIR T2:
cont.
Switches the stirrer at Tower 2
(only for 2-tower models)
cont., ON, OFF
(cont.= retain condition)
>manual stop
STIR MSB1:
cont.
Switches the stirrer at MSB connection 1
cont., ON, OFF
(cont. = retain condition)