7 connecting dosing devices and stirrers – Metrohm 789 Robotic Sample Processor XL User Manual
Page 38

2.2 Instrument setup
30
Metrohm Sample Processor, Installation
2.2.7 Connecting dosing devices and stirrers
Metrohm Sample Processors have various connections to which dosing
devices and stirrers can be connected.
Stirrers
The 722/802 Rod Stirrers and 741 Magnetic Stirrer have a DIN
connector and are connected to the rear panel of a tower (Tower stirrer
T1 and T2).
As with other Metrohm instruments, e. g. Titrandos, the three MSB
connectors (MSB = Metrohm Serial Bus) are very versatile. You can
connect an 801 Stirrer or an 804 Ti Stand to them.
Dosing devices
Up to three 700/800 Dosino or 685/805 Dosimat type dosing drives
can be operated by connecting them directly to the MSB sockets of a
Sample Processor. One 800 Dosino or 805 Dosimat can also be con-
nected to the MSB output connector of another instrument connected
to the MSB socket, such as the 801 Stirrer, in a so-called 'daisy chain'.
Please note that only one instrument of a particular type may be pre-
sent in such a daisy chain, e. g. a 801 Stirrer and a 805 Dosimat.
A 700 Dosino or a 685 Dosimat must be connected directly to the
MSB socket of the Sample Processor.
R S 2 3 2
M S B 1
M S B 3
M S B 2
R e m o te
Fig. 14 MSB connections
The instrument will only recognize automatically a connected dosing
device when it is switched on or after RESET.
Dosing device and
stirrer connections
MSB1…3