3.8 Dosing and liquid handling
106
Metrohm Sample Processor, Operation
3.8.8 Preparing the Dosing unit
The Dosing unit must first be prepared, i.e. the filling tubing and pipet-
ting tubing must be rinsed and completely filled. The cylinder contents
must then be ejected.
Command sequence:
MOVE 1 : ext.1
LIFT: 1 : work mm
DOS: 1.1 : prepar. mL
DOS: 1.1 : EndVol mL
DOS: 1.1 : compen. mL
11
Important!
Eject the cylinder con-
tents
This step should be carried out before every sample series involving pi-
petting. Before each pipetting process the Dosing unit cylinder contents
must be completely ejected. This is the only way to achieve reproduci-
ble, accurate pipetting.
The titration cell can then be aspirated, rinsed and prepared with fresh
solvent.
3.8.9 Pipetting
The pipetting procedure takes place in five steps:
• Ejection of cylinder contents (see above) and formation of separat-
ing bubble
• Move to sample
• Aspirate / Measure out sample
• Move to target
• Eject sample