32

Metrohm Sample Processor, Installation

Titrinos and Titrandos can be connected to a Sample Processor in the

same way and can also be controlled.

Control commands:

CTL:Rm : START Device1

starts

Titrino/Titrando

CTL:Rm : *************1

"

Final scan:

SCN:Rm : End1

waits for end of titration (EOD-impulse)

SCN:Rm : ****1***

"

SCN:Rm : Ready1

waits until Titrino/Titrando is ready

SCN:Rm : *******1

"

Sample Processor — pH / Ion Meter / Conductometer (692/712/713)

778
789

692

712
713


Control commands:

CTL:Rm : START device1

starts

instrument

CTL:Rm : *************1

"

CTL:Rm : METER Mode pH

switches to pH measurement

CTL:Rm : *********0001*

"

(not for 712)

CTL:Rm : METER Mode T

switches to temp. measurement

CTL:Rm : *********0010*

"

(not for 712)

CTL:Rm : METER Mode U

switches to mV measurement

CTL:Rm : *********0011*

"

(not for 712)

CTL:Rm : METER Mode I

switches to Ipol (mV measurement)

CTL:Rm : *********0100*

"

(not for 712)

CTL:Rm : METER Mode C

switches to Conc measurement

CTL:Rm : *********1000*

"

(for 692/781 only)

CTL:Rm : METER Cal pH

switches to pH calibration

CTL:Rm : *********0101*

"

(not for 712)

CTL:Rm : METER Cal C

switches to Conc calibration

CTL:Rm : *********1001

"

(for 692/781 only)

Cable 6.2141.020