2.3 Data transmission connections
32
Metrohm Sample Processor, Installation
Titrinos and Titrandos can be connected to a Sample Processor in the
same way and can also be controlled.
Control commands:
CTL:Rm : START Device1
starts
Titrino/Titrando
CTL:Rm : *************1
"
Final scan:
SCN:Rm : End1
waits for end of titration (EOD-impulse)
SCN:Rm : ****1***
"
SCN:Rm : Ready1
waits until Titrino/Titrando is ready
SCN:Rm : *******1
"
Sample Processor — pH / Ion Meter / Conductometer (692/712/713)
778
789
692
712
713
Control commands:
CTL:Rm : START device1
starts
instrument
CTL:Rm : *************1
"
CTL:Rm : METER Mode pH
switches to pH measurement
CTL:Rm : *********0001*
"
(not for 712)
CTL:Rm : METER Mode T
switches to temp. measurement
CTL:Rm : *********0010*
"
(not for 712)
CTL:Rm : METER Mode U
switches to mV measurement
CTL:Rm : *********0011*
"
(not for 712)
CTL:Rm : METER Mode I
switches to Ipol (mV measurement)
CTL:Rm : *********0100*
"
(not for 712)
CTL:Rm : METER Mode C
switches to Conc measurement
CTL:Rm : *********1000*
"
(for 692/781 only)
CTL:Rm : METER Cal pH
switches to pH calibration
CTL:Rm : *********0101*
"
(not for 712)
CTL:Rm : METER Cal C
switches to Conc calibration
CTL:Rm : *********1001
"
(for 692/781 only)
Cable 6.2141.020