2.3 Data transmission connections

34

Metrohm Sample Processor, Installation

Start commands:
In principle these are the same commands as those given in the previ-

ous section.

Final scan:
In principle these are the same commands as those given in the previ-

ous section, with the addition of:

SCN:Rm : EndMeter

waits for end impulse from 781

SCN:Rm : ***11***

"

Sample Processor— 2 x Titrino/Titrandos

B

A

Titrino

C

D

B

A

Titrino

C

D

778
789

Control commands:

CTL:Rm : START Device1

starts Titrino/Titrando 1

CTL:Rm : *************1

"

CTL:Rm : START Device2

starts Titrino/Titrando 2

CTL:Rm : ********1*****

"

CTL:Rm : START Device*

starts both Titrino/Titrandos

CTL:Rm : ********1****1

"


Final scan:

SCN:Rm : End1

waits for end of titration Titrino/Titrando 1

SCN:Rm : ****1***

"

SCN:Rm : End2

waits for end of titration Titrino/Titrando 2

SCN:Rm : *1******

"

SCN:Rm : Ready1

waits until Titrino/Titrando 1 is ready

SCN:Rm : *******1

"

SCN:Rm : Ready2

waits until Titrino/Titrando 2 is ready

SCN:Rm : **1*****

"

SCN:Rm : Ready*

waits until both Titrino/Titrandos are ready

SCN:Rm : **1****1

"

The combined operation of the Titrino/Titrando/pH meter can also be

carried out without any problems by the same means.

Device 1

Device 2

Cable 6.2141.030