Metrohm Sample Processor, Annex

131

20 DOS: 1.1 : 5.4 mL

21 WAIT: pause 3 s

22 DOS: 1.2 : fill mL

23 DOS: 1.1 : endVol mL

24 DOS: 1.1 : compen. mL

25 LIFT: 1 : shift mm

26 WAIT: pause 5 s

27 CTL:Rm: START device1

28 SCN:Rm : Ready1

29 PUMP 1.* : 30 s

30 STIR: MSB1 : OFF s

>final sequence

1 MOVE 1 : ext.1

2 LIFT: 1 : work mm

3 DOS: 1.1 : empty mL

4 LIFT: 1 : shift mm

5 PUMP 1.1 : 15 s

>changer settings

...

>stirring rates

>Dosing unit def.

dosing drive 1

dos.rate 10.0 ml/min

fill.rate 10.0 ml/min

dosing port 1

dosing 2 port 3

filling port 2

rinsing port 2

preparation port 1

drain port 4

cock direction: not over

not over port 4

>timeout settings

SCAN timeout: OFF min

on SCAN timeout: error

>manual stop

CTL Rmt: STOP device*

CTL RS232:

PUMP: OFF

STIR T1: cont.

STIR T2: cont.

STIR MSB1: cont.

STIR MSB2: cont.

STIR MSB3: cont.

------------

- Eject sample (+ air-gap bubble )

- Wait for drips

- Refill solvent in dosing cylinder

- Eject solvent into titration cell (rinse)

- Eliminate slack

- Run lift to shifting height

- Wait 5 seconds

- Start titrator

- Wait for end of titration

- Rinse and aspirate titration cell

- Switch off stirrer

- Move robotic arm to titration cell

- Run lift to working height

- Empty dosing cylinder and tubing

- Run lift to shifting height

- Rinse and aspirate titration cell
Changer settings

Stirring rates
Settings for dosing device

- Reduce dosing rate

- Reduce filling rate

- Do not rotate to open Port 4 (in bottle)

- see above

Actions after manual stop
- Stop titrator

- Switch off pump