Page 87

background image

3.4

Commands

Metrohm Sample Processor, Operation

79

The DOS command is used for controlling Dosimats and Dosinos. Up

to 3 Dosinos or Dosimats can be addressed individually or simultane-

ously via the MSB bus control.

The 1st parameter selects the dosing device and the corresponding

dosing port, at which the required function is to be carried out. If '*' is

entered then the default port for the corresponding function will be used

(e.g. dosing port = 1, filling port = 2, etc.)

Please note

After dosing the Sample Processor

will not automatically fill the

dosing cylinder. If required, this can be programmed with the com-

mand

DOS: 1.* : fill

, see below.

It is possible to enter the volume to be added directly as the 2

nd

pa-

rameter or to carry out specific functions of a Dosino. Negative volumes

can also be added, i.e. a certain volume can be aspirated and then

ejected again (pipetting). The minus sign is entered with the <*> key.

The functions listed below are used for complex liquid handling tasks

such as pipetting.

Dosing functions:

fill

Fills the Dosimat and Dosino cylinder.

release

Prepares the Dosimat or Dosino for changing the Ex-

change/Dosing unit. The dosing cylinder is filled via the

rinse port. The stopcock is turned to the exchange posi-

tion (Port 2).

prepar.

Preparation cycle (PREP) for Dosinos. All tubing is rinsed

and filled completely.

empty

Tubing system and Dosino cylinder are emptied.

eject

Complete Dosino cylinder contents are ejected.

endVol

Ejects the cylinder content to the max. volume mark.

compen.

Cancels the mechanical play between the dosing piston

and spindle.

In the parameter menu under

>Dosing unit def.

the port assignments

of the Dosinos as well as the dosing and filling rates can be defined for

specific methods. This can also be carried out in a sequence with the

appropriate DEF command.

More details about Dosino commands are given on p. 101ff.

The Sample Processor automatically recognizes whether a Dosimat or

a Dosino is connected.