Metrohm 789 Robotic Sample Processor XL User Manual

Page 70

background image

3.2 Configuration

62

Metrohm Sample Processor, Operation

>recall rack

shift position T2

0 mm

Shifting position for sample positions at
Tower 2 *

)

0…325 mm

(in mm from upper stop point)

>recall rack

special pos. T1

0 mm

Special position for sample positions at
Tower 1 *

)

0…325 mm

(in mm from upper stop point)

>recall rack

special pos. T2

0 mm

Special position for sample positions at
Tower 2 *

)

0…325 mm

(in mm from upper stop point)

>recall rack

beaker radius

* mm

Effective beaker radius for the sample
positions on the rack

*,

1…100 mm

* = any value

This setting is required for an automatic check of the beaker radius, see
p. 58.

>recall rack

beaker sensor:

Tower

Selection of the beaker sensor

Tower,

SwingH, OFF


The beaker sensor checks the presence of a sample beaker when executing

a MOVE command.

>recall rack

rack offset

0.00°

Correction for the rotation angle of a

sample rack *

)

-5.00…0.00…5.00°

*

)

All the above lift positions and the rack offset can be set by using the

function.

Special beaker positions

>>recall rack

>>>special positions

Submenu special beaker positions

Open with


Special beaker positions are predefined places on a sample rack that are not
treated as sample positions. They can be occupied by rinsing beakers or
conditioning beakers and selectively addressed during a run sequence. Up
to 16 special beaker positions can be defined per rack. For each spezial
beaker the work position of the lift (tower 1 and 2) and the beaker radius can
be set, see above.

Selection of special beaker

>>>special positions

special beaker

1

1…16

Use

for next higher

level