3 data transmission connections, 1 remote connections – Metrohm 789 Robotic Sample Processor XL User Manual
Page 39

2.3 Data transmission connections
Metrohm Sample Processor, Installation
31
2.3 Data transmission connections
Connection cables
Only Metrohm cables should be used to connect a Metrohm Sample
Processor to other instruments. Only these will guarantee interference-
free data transmission.
Note:
The Remote cables for Metrohm Sample Processors and Sample
Processors have markings at each end of the cable to indicate the in-
strument that the particular connector is intended to be used with and
the connection into which it is to be plugged.
Example:
Titrino B
692 / 712 / 713
Fig. 15 Remote cable
The Sample Processor must be switched off before peripheral devices
are connected, as otherwise the instruments could be damaged.
2.3.1 Remote
connections
Sample Processor — Titrino / Titrando
with standard cable (and 6.2148.010 Remote Box)
B
A
Titrino
C
D
778
789
MSB
778
789
Titrando
6.2148.010
In order to connect Remote cables to Titrandos and the 780/781
pH/Ion Meter an 8.2148.010 Remote Box is required as an additional
adapter; this is connected to one of the MSB sockets (MSB1…4), see
above.
Cable 6.2141.020
Cable 6.2141.020