2.3 Data transmission connections
Metrohm Sample Processor, Installation
31
2.3 Data transmission connections
Connection cables
Only Metrohm cables should be used to connect a Metrohm Sample
Processor to other instruments. Only these will guarantee interference-
free data transmission.
Note:
The Remote cables for Metrohm Sample Processors and Sample
Processors have markings at each end of the cable to indicate the in-
strument that the particular connector is intended to be used with and
the connection into which it is to be plugged.
Example:
Titrino B
692 / 712 / 713
Fig. 15 Remote cable
The Sample Processor must be switched off before peripheral devices
are connected, as otherwise the instruments could be damaged.
2.3.1 Remote
connections
Sample Processor — Titrino / Titrando
with standard cable (and 6.2148.010 Remote Box)
B
A
Titrino
C
D
778
789
MSB
778
789
Titrando
6.2148.010
In order to connect Remote cables to Titrandos and the 780/781
pH/Ion Meter an 8.2148.010 Remote Box is required as an additional
adapter; this is connected to one of the MSB sockets (MSB1…4), see
above.
Cable 6.2141.020
Cable 6.2141.020