3 composition of a method, 1 run sequences and method parameters – Metrohm 789 Robotic Sample Processor XL User Manual

Page 74

background image

3.3 Composition of a method

66

Metrohm Sample Processor, Operation

3.3 Composition of a method

All the settings of the parameter menu form a method and can be

stored as such.

3.3.1 Run sequences and method parameters

PARAM

Main menu:

parameters

number of samples:

rack

Number of samples to be processed

1…999,

rack,∗

rack = one sample rack rotation
∗ =

infinite

In the

rack

setting all the sample positions of the current rack will be

processed (max. no. of rack positions minus number of special beak-

ers), with only the positions occupied by sample beakers being
counted. It is important that the Sample Processor can recognize the
rack. This is only possible when the rack is in initial position. We rec-
ommend that each time a rack is changed the Sample Processor is

initialized with .

parameters

>start sequence

Open submenu with

parameters

>sample sequence

Use < > or < > to move up and down by one menu
item

parameters

>final sequence

Use or to reach the first and last

menu item

parameters

>changer settings

Use to return to normal operating condition

parameters

>stirring rate


parameters

>Dosing unit def.

parameters

>timeout settings


parameters

>nanual stop



Submenus:

In each of the submenus

>start sequence

,

>sample sequence

and

>final sequence

up to 99

command lines can be entered as a run sequence. The commands can be entered directly via

the keypad. The command keys on the right-hand side of the keypad are available.

parameters

>start sequence

Commands for the start sequence of the sample series

Open the submenu with