Metrohm Sample Processor, Annex

>sample sequence

1 MOVE 1 : sample

2 LIFT: 1 : work mm

3 CTL:Rm: METER mode C

4 CTL:Rm: START device1

5 SCN:Rm : EndMeter

6 LIFT: 1 : rinse mm

7 PUMP 1.1 : 5 s

8 WAIT: pause 3 s

>final sequence

1 MOVE 1 : spec.1

2 LIFT: 1 : work mm

>changer settings

rack name: *

lift rate T1 25 mm/s

lift rate T2 25 mm/s

shift rate 20°/s

shift direction: auto

swing rate T1 55°/s

swing rate T2 55°/s

on beaker error: MOVE

>stirring rates

>Dosing unit def.

>timeout settings

SCAN timeout: OFF min

on SCAN timeout: error

>manual stop

CTL Rmt: STOP device*

CTL RS232:

PUMP: OFF

STIR T1: cont.

STIR T2: cont.

STIR MSB1: cont.

STIR MSB2: cont.

STIR MSB3: cont.

------------

- Move to sample

- Run lift to work height

- Switch ion meter to concentration measurement

- Start measurement

- Wait for measurement

- Runlift to rinse height

- Rinse for 5 seconds with pump

- Allow to drip for 3 seconds

- Move to rinse beaker

- runlift to work height
Changer settings

Stirring rates
Settings for dosing device

Actions after manual stop
- Stop ion meter