Metrohm 789 Robotic Sample Processor XL User Manual
Page 107

3.8 Dosing and liquid handling
Metrohm Sample Processor, Operation
99
Eject
DOS: X.Y : eject mL
Ejects the complete dosing cylinder con-
tents
The cylinder are completely ejected via the given port. The piston
moves down to the stop. If * is entered as a wild card then the port
given under
>dosing drive X
dosing port Y
will be used (default: Port 1).
Endvolume
DOS: X.Y : endVol mL
Ejects the cylinder volume to the max. vol-
ume mark
The content of the dosing cylinder is ejected via the given port. The pis-
ton is run the nominal volume. If * is entered as a wild card then the
port given under
>dosing drive X
dosing port Y
will be used (default: Port 1).
Compensate
DOS: X.Y : compen. mL
Eliminates the mechanical play
The mechanical play between the dosing piston and drive spindle is
eliminated after the stopcock has been rotated to the given port. If * is
entered as a wild card then the port given under
>dosing drive X
dosing port Y
will be used (default: Port 1).