3.8 Dosing and liquid handling

Metrohm Sample Processor, Operation

99

Eject

DOS: X.Y : eject mL

Ejects the complete dosing cylinder con-

tents

The cylinder are completely ejected via the given port. The piston

moves down to the stop. If * is entered as a wild card then the port

given under

>dosing drive X

dosing port Y

will be used (default: Port 1).

Endvolume

DOS: X.Y : endVol mL

Ejects the cylinder volume to the max. vol-

ume mark

The content of the dosing cylinder is ejected via the given port. The pis-

ton is run the nominal volume. If * is entered as a wild card then the

port given under

>dosing drive X

dosing port Y

will be used (default: Port 1).

Compensate

DOS: X.Y : compen. mL

Eliminates the mechanical play

The mechanical play between the dosing piston and drive spindle is

eliminated after the stopcock has been rotated to the given port. If * is

entered as a wild card then the port given under

>dosing drive X

dosing port Y

will be used (default: Port 1).