5 rs232 settings – Metrohm 789 Robotic Sample Processor XL User Manual
Page 73

3.2
Configuration
Metrohm Sample Processor, Operation
65
3.2.5 RS232
settings
The correct settings of the interface parameters of the serial RS232 interface is essential for the
perfect functioning of data transmission to and from the Sample Processor. This includes the
control of the instrument by using a PC software such as TiNet or Tiamo, backing up of con-
figurations or methods with the VESUV software from Metrohm or printing out a parameter re-
port on a connected printer.
configuration
>RS232 settings
Submenu for serial interface settings
Open the submenu with
>RS232 settings
baud rate:
9600
Transmission rate
in baud
300,600,1200,
2400,4800,9600, 19200
>RS232 settings
data bit:
8
Number of data bits
7,8
>RS232 settings
stop bit:
1
Number of stop bits
1,2
>RS232 settings
parity: none
Parity selection
even, odd,
none
>RS232 settings
handshake: HWs
Handshake selection
HWs,SWchar,SWline,
none
>RS232 settings
character set:
IBM
Character set for printer and PC
(printer emulation)
IBM,HP,Epson,
Seiko,Citizen
The settings for the printers recommended by Metrohm are listed on p. 37ff.
If your printer is not shown there then we recommend the setting "Epson". It
may be necessary to consult the printer manual. For data transmission using
personal computers you should select "IBM".
>RS232 settings
RS control:
ON
Switch data reception on/off
ON,OFF
If Remote control is switched off then no data will be received, but reports
can still be printed out.
Use
for next higher
level