4.3 Systems
790 Personal IC
85
INJECT done
Injection valve has been switched to the
"INJECT" position.
Finished
Determination has been finished.
Executing *.smt [*.que]
The sample queue
*.que
is processed using the
system file
*.smt
.
SHUTDOWN
790 Personal IC hardware is shutdown.
Error messages
Detection of hard-
ware failed
Bad connection between PC and 790 Personal
IC or instrument switched off (check connecting
cable or switch on instrument).
E1
Program checksum wrong (call Metrohm
service).
E2
RAM faulty (call Metrohm service).
E200
Invalid instrument adjustment (call Metrohm
service).
E237
Storage of configuration values failed (repeat last
action; if error reappears, call Metrohm service).
E238
Storage of instrument number failed (repeat last
action; if error reappears, call Metrohm service).
E240
EEPROM faulty (call Metrohm service).
E258
Storage of setup values failed (repeat last action;
if error reappears, call Metrohm service).
E295
Storage of memory values failed (repeat last
action; if error reappears, call Metrohm service).
E296
Instrument stopped (restart instrument; if error
reappears, call Metrohm service).
E298
Storage of flow correction value failed (repeat
last action; if error reappears, call Metrohm
service).
E299
Storage of break time values failed (repeat last
action; if error reappears, call Metrohm service).
E300
High-pressure pump faulty (restart pump; if error
reappears, call Metrohm service).
E301
Injection valve blocked (check injection valve; if
error reappears, call Metrohm service).
E302
Suppressor module blocked (check suppressor;
if error reappears, call Metrohm service).
E303
Storage of maintenance information failed
(repeat last action; if error reappears, call
Metrohm service).