4 Operation
790 Personal IC
120
Mode
The two options
Overwrite
(overwrite the file) or
Append
(append to existing file) are available.
Character set
The two options
Windows
(ANSI) or
DOS
(ASCII)
are available. These settings are essential for
printing of non-English symbols (e.g. ц, д, ь).
Custom program
Path and name of the program to be started af-
ter report output. This option enables to transfer
the report to a database, an electronic table or
another application for further processing.
Report elements
If
Quantification method = Custom
has been selected, the peak table can
be customized on the user's choice. After clicking the
but-
ton, the following columns can be included in the peak table.
number
Peak
number.
retention time
Retention time of the component (in minutes, ir-
respective of the chosen retention units on the
chromatogram graph axes). The total value in
the column is equal to the chromatogram dura-
tion.
halfwidth
Width of the peak at half height (in minutes).
height
Height of the peak (in µS/cm). The total value in
this column is the sum of heights for all identi-
fied peaks.
height%
Normalization of peak heights for all peaks us-
ing the NORM value entered in the
Total % for
normalization
field (default setting 100%):
H
i
% = NORM
•
H
i
/
H
i
area
Area of the peak. Depends on the units on the
X and Y axes of the chromatogram. The total
value in this column is the sum of areas for all
identified peaks (including universal compo-
nent).
area%
Normalization of peak areas for all peaks using
the NORM value entered in the
Total % for nor-
malization
field (default setting 100%):
A
i
% = NORM
•
A
i
/
A
i
capacity factor
The capacity factor k'
i
of the component is
equal to the ratio of its corrected retention time
(t - t
0
)
to the void time of the system t
0
:
k'
i
= (t
i
- t
0
) / t
0
Total for this column is equal to the capacity
factor of the last peak of the chromatogram.