4 Operation
790 Personal IC
140
Level
Calibration level (see section 4.4.5) for the
sample: Level
0
stays for normal analysis, levels
1
and greater for calibration runs. Correct filling
of this column enables to calculate calibration
coefficients automatically.
Volume
Injected volume of sample in µL.
Amount
Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
=
C
c
•
Amount
s
/
Amount
c
Shut down system after the queue finishes
If this option is enabled, the system hardware is
shut down after the sample queue has been fin-
ished (all pumps are switched off), but it re-
mains connected.
Close queue dialog window after the queue finishes
If this option is checked, the sample queue
window is closed automatically after the sample
queue has been finished.
Start execution of the sample queue from the
first line with
Started = 0
.
Stop execution of the sample queue after the
current determination has been finished.
Open the sample queue editor program for
editing the sample queue table (see section
4.6.3).
Close the sample queue overview table.
Start sample queue
SAMPLE QUEUE OVERVIEW /
Control / Start
Start execution of the sample queue from the first row with
Started = 0
.
For each row, the sample-specific data are transferred to the method of
the current determination and the parameter
Started
is set to
1
for this
row.
If processing the sample queue is interrupted with
then it can
be continued again with
(with a running chromatogram only
when data acquisition is finished).