4 Operation

790 Personal IC

140

Level

Calibration level (see section 4.4.5) for the

sample: Level

0

stays for normal analysis, levels

1

and greater for calibration runs. Correct filling

of this column enables to calculate calibration

coefficients automatically.

Volume

Injected volume of sample in µL.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

=

C

c

Amount

s

/

Amount

c

Shut down system after the queue finishes

If this option is enabled, the system hardware is

shut down after the sample queue has been fin-

ished (all pumps are switched off), but it re-

mains connected.

Close queue dialog window after the queue finishes

If this option is checked, the sample queue

window is closed automatically after the sample

queue has been finished.

Start execution of the sample queue from the

first line with

Started = 0

.

Stop execution of the sample queue after the

current determination has been finished.

Open the sample queue editor program for

editing the sample queue table (see section

4.6.3).

Close the sample queue overview table.




Start sample queue

SAMPLE QUEUE OVERVIEW /

Control / Start

Start execution of the sample queue from the first row with

Started = 0

.

For each row, the sample-specific data are transferred to the method of

the current determination and the parameter

Started

is set to

1

for this

row.

If processing the sample queue is interrupted with

then it can

be continued again with

(with a running chromatogram only

when data acquisition is finished).