5.2 Maintenance and servicing

790 Personal IC

161

6 Mounting the pump head

• Reassemble the components of the pump head 34

34

34

34 as shown

in Fig. 17. Tighten the screw 84

84

84

84 by hand. First screw in piston

cartridge 88

88

88

88 manually until the stop is reached and then use

a wrench to turn it through a further

15°. Firmly tighten the two

valve screw holders 95

95

95

95 with a wrench.

• Reattach pump head 34

34

34

34 to the pump with the help of the 4

fixing screws 33

33

33

33. Firmly tighten them with the 6.2621.030 Al-

len key.

To ensure that the pump head is not positioned wrongly, the holes at

the rear for the clamping bolts have different depths, i.e. 1 clamping

bolt is longer than the rest. The deepest hole must naturally accom-

modate the longest bolt. If this is not the case, the pump will not

function properly.

• Screw connection capillary 35

35

35

35 to pump head 34

34

34

34 (see Fig. 3

and Fig. 4).

• Connect aspirating tubing to aspirating capillary 32

32

32

32 at the

pump head 34

34

34

34.


Inlet valve

Outlet valve

6.2824.090 6.2824.080

94

94

94

94

99

99

99

99

100

100

100

100

101

101

101

101

102

102

102

102

103

103

103

103

104

104

104

104

98

98

98

98

99

99

99

99

100

100

100

100

103

103

103

103

102

102

102

102

101

101

101

101

104

104

104

104

Fig. 19: Components of inlet valve

93 and

outlet valve

94

98

98

98

98 Valve housing

99

99

99

99 Sealing ring (black)

100

100

100

100 Sleeve

101

101

101

101 Sapphire sleeve

The bright side must point towards

the sapphire sphere

102

102

102

102 Sapphire sphere

103

103

103

103 Ceramic holder for sapphire

sphere

The large recess must point in the

direction of the sapphire sphere

104

104

104

104 Seal

The larger opening must point

outwards