6.1 Technical data

790 Personal IC

173

6.1.6 Suppressor

module

Switching duration

140 ms

Pressure resistance

2.5 MPa (25 bar)

6.1.7 RS232

interface

Connector

Dsub 9 pin (male)

Function

TxD and RxD signal for connection with software

handshake

Default settings

9600 baud, 8 bit, 1 stop bit, no parity, XON/XOFF

Pin assignment

6.1.8 Remote

interface

Connector

Dsub 25 pin (female)

Circuit diagram for

output line 3

Potentials



Pin 3

Remote

Pin 25

Pin 14

Pin 15

0V

+5V