beautypg.com

3 eluents, 4 peristaltic pump – Metrohm 790 Personal IC User Manual

Page 162

background image

5.1 Practical notes on ion chromatography

790 Personal IC

153

5.1.3 Eluents

Treatment

For the preparation of the eluents, only chemicals of a purity degree of

at least "

p.a." should be used. For diluting, please use only high purity

water.
Fresh eluents should always be

microfiltered (0.45 µm filter) and de-

gassed (with N

2

, He or vacuum). For alkaline eluents and eluents with

low buffering capacity one should preferably use a

CO

2

absorber.

The supply vessel containing the eluent must be closed as tightly as

possible to avoid excessive evaporation. This is primarily important with

eluents containing organic solvents (e.g. acetone), the evaporation of

which can lead to drifts in the long term. If work is performed in a very

sensitive range, even if one drop of condensate falls back in the eluent

this can cause a noticeable change in the background conductivity.

Influence of various parameters on anion columns

Concentration:

An increase in the concentration usually leads

to shorter retention times and quicker separa-

tion, but also to a higher background conduc-

tivity.

pH:

pH alterations lead to shifts in the dissociation

equilibrium and thus to changes in the reten-

tion times.

Organic modifiers:

Addition of an organic solvent (e.g. methanol,

acetone, acetonitrile) to aqueous eluents ge-

nerally accelerates lipophilic ions.

Eluent change

When the eluent is changed, it must be ensured that

no precipitates

can be formed. Solutions used in direct succession must therefore be

miscible. If the system has to be rinsed with an organic solution, several

solvents with increasing or decreasing lipophilic character may possibly

have to be used (e.g. water ↔ acetone ↔ chloroform).

5.1.4 Peristaltic

pump

The 6.1826.060 pump tubings used by the peristaltic pump are con-

sumable material with a limited lifetime and should be exchanged at

regular intervals (approx. every 4 weeks under continuous use; see

section 5.2.10).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in section 2.9.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges 40

40

40

40 should

be raised completely by loosening the snap-action lever 43

43

43

43 on the

right-hand side (the set contact pressure remains unchanged).