2.6 High-pressure pump

790 Personal IC

29

• Close the rotary knob on purge valve 31

31

31

31 by turning it in a

clockwise direction (see Fig. 3 and Fig. 4).

• Remove the syringe from coupling 24

24

24

24.

5 Rinse pulsation dampener (if present)

• Place a beaker beneath the column connection capillary 20

20

20

20.

• Click the

button for

IC pump

to switch on the high-

pressure pump and rinse the pulsation dampener 52

52

52

52 filled

with isopropanol for ca. 10 min with eluent.

• Click the

button for

IC pump

to switch off the high-

pressure pump.

6 Reduce flow rate

• Reset the original flow rate under

Flow

(e.g.

0.5 mL/min

).

• Click

to send this value to the 790 Personal IC.