6.1 Technical data

790 Personal IC

171

6 Appendix

6.1 Technical

data

6.1.1 Conductivity

measurement

Measurement range

0…1000 µS/cm (resolution: 0.56 nS/cm)

Maximum error

± 10 µS/cm and

± 1 % of measurement value (k = 16.7/cm)

Linearity

Deviations < ± 5 µS/cm

Noise

typ. 3 nS/cm

Drift (electronic)

typ. < 10 nS/cm / h

Temperature dependence

typ. < 40 nS/cm / °C

Reserve range

> 33 % (k = 16.7/cm)

Sampling rate

10 measurements/s (fixed)

6.1.2 Conductivity

detector

Construction

Thermostatted conductivity detector with

2 ring-shaped steel electrodes

Measurement principle

Alternating current measurement with 1 kHz

frequency and ca. 1.7 V amplitude (peak to

peak).

Cell volume

1.5 µL

Cell constant

approx. 17 /cm (the exact value is printed on the

detector)

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to operating

temperature

Operating temperature

40 °C

Max. temperature deviation

± 2.5°C

Heating time

≥ 30 min

Temperature stability

≤ 0.01°C at constant ambient temperature

Connection for detector block Dsub 15 pin (female)

6.1.3 Injection

valve

Actuator switching duration

100…150 ms

Pressure resistance

25 MPa (250 bar)