background image

4 Operation

790 Personal IC

114

Comment

User defined comment from the

Comment

tab of

the passport.

COMMENT
Beispielmethode zur Wasserbestimmung.

GLP Log

Method and Data GLP messages from the

Method Log

and

Data Log

tabs of the passport.

METHOD GLP LOG
13/04/2000 08:32:49 W. Terzer

04/05/2000 10:45:23
790 Personal IC 1.0 Apr 27 2000 14:05:01

DATA GLP LOG
Analysing data of channel :Cond
Time constant, ms: 100.00
Sigma: 0.0
Rejection difference: 1.0

11/04/2000 18:01:45
Status: Processing
Data saved to disk file c:\programme\ic 790\data\K4111745.CHW

More items to report
This part of the

Report options

window contains a list of rarely used re-

port items that can be included into the report on the user's choice:

Acquisition

Parameters used for data acquisition.

ACQUISITION PARAMETERS
Method duration: 16.00 min
Run duration: 16.01 min
Measurements (method): 9606
Measurements (run): 9606
Freq.divisor: 1
Sampling: 10.00 pts/sec
Device: 790 Personal IC

Integration

Integration parameters and integration events.

INTEGRATION DEFAULTS
Delay: 3.60 min
Width: 5.00 sec
Slope: 20.00
MinHeight: 0.01

No. min
1 0.00 Enable valley-to-valley
2 6.00 Disable valley-to-valley
3 9.50 Set peak end