Metrohm 790 Personal IC User Manual
Page 125
4 Operation
790 Personal IC
116
Channel table
Parameters of the data acquisition channel
used.
CHANNELS TABLE
No Name Units Input Minimum Zero Maximum Range Coefficient Noise Shift
1 Cond uS/cm 1 -8388607 0 8388607 8.389e+01 1.000e-05 1.52 0
# Name/Units Noise RMS PeakToPeak Drift/hour
1 Cond 1.5 52871.6 512493.0 4232.464
uS/cm 1.52e-05 0.529 5.12 0.0423
Name
Channel’s name. This text appears as
Channel
label
in the chromatogram.
Units
Units of detector response (µS/cm).
Input
Number of data acquisition channel.
Minimum
Minimum value of the linear range of the AD
converter (in ADC conversion units). This pa-
rameter is used to detect an underflow condi-
tion.
Zero
A signal level on a baseline (in ADC conversion
units). This parameter is a result of the ADC
calibration.
Maximum
Maximum value of the linear range of the AD in-
terface (in ADC conversion units). This parame-
ter is used to detect overflow condition.
Range
An input signal value in maximum in µS/cm.
Range
= (
Maximum
-
Zero
)
•
Coef
Coef
ADC sensitivity coefficient (weight of ADC bit in
µS/cm).
Noise
Estimated baseline noise value of the channel
in ADC conversion units (bits).
Shift
Not used for 790 Personal IC.
Report destination
The following output devices can be selected simultaneously in any
combination as targets for report output:
Screen
Output report (without curves) to screen.
Printer
Output report to printer.
File