3 method setup – Metrohm 790 Personal IC User Manual
Page 102
![background image](/manuals/586529/102/background.png)
4.4 Methods
790 Personal IC
93
4.4.3 Method
setup
790 PERSONAL IC / Method / Method setup
This menu item opens the
Method setup
window that includes the most
common parameters for data acquisition and evaluation of the method
and consists of the following tabs:
General
General information on determination.
Processing
Set actions that are performed when the chroma-
togram finishes.
Math
Parameters that are used for various types of cal-
culations.
General
Tab
General
of
Method setup
window with general description of the
method and determination. This tab is identical to the
General
tab called
with
Method / Passport
(see section 4.4.2).
Processing
Tab
Processing
of
Method setup
window for definition of actions that are
performed during the run or when the chromatogram finishes. Auto-
matic post-run data processing is especially useful for chromatographic
systems that include an autosampler.
Save chromatogram after the run
If checked, autosaving of the chromatogram
will be performed on data acquisition ending.
while running every ... min
Save the chromatogram during the run for data
safety at the time interval set (if
0
is set, the
chromatogram is not saved during the run).
Show all
If checked, scales on X and Y axes will be
automatically chosen so that the recorded
chromatogram fits the window.