Metrohm 790 Personal IC User Manual
Page 173
5 Notes – Maintenance – Faults
790 Personal IC
164
2 Dismantle suppressor
• Unscrew nut 105
105
105
105 from suppressor holder 108
108
108
108.
• Pull out connection piece 106
106
106
106 and suppressor rotor 107
107
107
107 from
suppressor holder 108
108
108
108 (the connection piece and the rotor
normally stick together).
• Loosen connection piece 106
106
106
106 from suppressor rotor 107
107
107
107.
3 Clean input and output leads
• Connect each of the 6 capillary tubings attached to connec-
tion piece 106
106
106
106 to the high-pressure pump one after another,
and pump through ultrapure water.
• Check whether solution emerges from connection piece 106
106
106
106.
If one of the input or output leads remains blocked, replace
the connection piece 106
106
106
106 (order number 6.2832.010).
4 Clean suppressor rotor
• Clean the sealing surface of suppressor rotor 107
107
107
107 using a
lint-free cloth and ethanol.
5 Insert suppressor rotor
• Insert suppressor rotor 107
107
107
107 in suppressor holder 108
108
108
108 in such
a way that the tubing connections at the rear of the rotor fit in
the corresponding openings inside the rotor, and that one of
the three holes in the rotor can be seen from below in one of
the openings of the holder.
• If the rotor has been inserted correctly, its sealing surface will
be about 4 mm inside the holder. If this is not the case, bring
the rotor into the correct position from below with the aid of a
sharp object (e.g. a screwdriver).
6 Clean connection piece
• Clean the sealing surface of connection piece 106
106
106
106 using a
lint-free cloth and ethanol.
7 Insert connection piece
• Insert connection piece 106
106
106
106 in suppressor holder 108
108
108
108 in such
a way that connection "1" is at the top, and that the three lugs
on the connection piece fit in the corresponding openings of
the holder.
• Screw nut 105
105
105
105 onto the thread of suppressor holder 108
108
108
108
manually (do not use tools).
8 Connect and condition the suppressor
• Reconnect the suppressor to the IC system.
• Before switching the suppressor to the next position for the
first time, rinse all 3 suppressor units with solution for 5 min.