beautypg.com

3 recycling, 4 shutdown, 5 changing separating columns – Metrohm 790 Personal IC User Manual

Page 165

background image

5 Notes – Maintenance – Faults

790 Personal IC

156

5.2.3 Recycling

To keep the eluent consumption between injections to a minimum

when the system is at rest (e.g. overnight), the so-called recycling pro-

cedure can be used. In recycling the eluent exiting the outlet capillary of

the detector block is led back directly to the eluent supply vessel. The

IC system is thus quickly ready for new injections without a long condi-

tioning period.

The recycling procedure must not be used

in operation with the suppressor module,
with alkaline eluents,
with the 6.1010.000 IC Cation column METROSEP Cation 1-2.

5.2.4 Shutdown

If the 790 Personal IC is shut down for a considerable length of time,

the entire IC system (

without column and suppressor) must be rinsed

free from salt with methanol/water (1:4) to avoid crystallization of elu-

ent salts with the corresponding subsequent damage.
For rinsing, the connections to the separating column and the suppres-

sor module are removed; the two capillaries 20

20

20

20 and 37

37

37

37 (see Fig. 13 and

Fig. 15) are directly connected to each other using coupling 26

26

26

26

(6.2620.060). Rinse with methanol/water (1:4) until the conductivity

drops below 10 µS/cm.

5.2.5

Changing separating columns

Identical separation system

If you wish to replace an IC separating column by a column of the same

type, proceed as follows (see Fig. 13 and Fig. 15):

1 Remove old column

• Switch off high-pressure pump and wait for pressure to drop.
• Unscrew column 63

63

63

63 from inlet capillary 37

37

37

37 of the detector

block or from suppressor inlet capillary 78

78

78

78.

• Unscrew column 63

63

63

63 from column connection capillary 20

20

20

20 or

the precolumn.

2 Connect new column to injector

• Remove end caps from column 63

63

63

63.

• Screw inlet end of separating column 63

63

63

63 (note flow direction)

to column connection capillary 20

20

20

20 or to the precolumn (see

section 2.7.7/section 2.7.8).