2.10 Connection of external devices
2.10.1 Connection of the 750 Autosampler
The 750 Autosampler available from Metrohm as an option is an auto-
matic sampler for ion chromatography. It accommodates max. 128
samples each of 730 µL, which are automatically transferred to the
sample loop attached to the injection valve of the 790 Personal IC. The
samples on the 750 Autosampler are changed by a remote signal that
is outputted by the 790 Personal IC (790 Personal IC as "Master").
In order to connect the 750 Autosampler the 6.2128.160 Cable that is
available as an option is additionally required. The 750 Autosampler is
connected to the 790 Personal IC as follows:
1 Electrical connection 790 – 750
• Connect 6.2140.010 Connector plug to contact closure strip
10
10
10
10 of the 750 Autosampler (see 750 Instructions for Use).
• Connect one end of the 6.2128.160 Cable to the connections
13 "EXTERNAL INJECT INPUT" and 14 "GND" on the connec-
tor plug. The two cable ends are appropriately inscribed with
"750" (white) and "GND" (brown).
• Connect the other end of the 6.2128.160 Cable to the remote
interface 16
16
16
16 of the 790 Personal IC (see Fig. 2).
2 Tubing connection 750 – injection valve
• Pull aspiration tubing 23
23
23
23 completely out from feedthrough 4
4
4
4
and screw it off from connection "1" of injection valve 22
22
22
22.
• Cut transfer tubing 8888 mounted on the 750 Autosampler (see
750 Instructions for Use) to the shortest possible length be-
tween the needle of the 750 Autosampler and injection valve
22
22
22
22 of the 790 Personal IC.
• Pull the free end of transfer tubing 8888 through the feedthrough
4
4
4
4 on the 790 Personal IC and screw it onto connection "1" of
injection valve 22
22
22
22 using a 6.2744.010 PEEK compression fit-
ting.
3 Tubing connection injection valve – waste
• Remove coupling 24
24
24
24 and 6.2744.010 Compression fitting
from syringe tubing 25
25
25
25.
• Lead free end of syringe tubing 25
25
25
25 to the waste container and
fix in place.
4 Settings at the 750 Autosampler
• Set parameter "
Run time
" to
0.0
to activate the external start
of the 750 Autosampler (see section 3.5 of the 750 Instruc-
tions for Use). Filling the sample loop of the injection valve
only takes place when the impulse produced by the 790 Per-