2 sample queue control – Metrohm 790 Personal IC User Manual
Page 148
4.6 Sample queue
790 Personal IC
139
4.6.2
Sample queue control
Sample queue overview table
SYSTEM / System / Sample queue
This menu item opens the
OPEN
window for opening an existing sample
queue file
*.que
or creation of a new sample queue file
*.que
by entering
a new name. After confirmation with
the sample queue overview
table is opened showing the current status of the sample queue:
No
Row
number.
System
System file to be used for the determination.
Ident
User defined identifier for chromatogram. Will
be placed into appropriate passport field when
starting the chromatogram.
Vial
Autosampler vial position to take sample from.
This value will not be transferred automatically
to the autosampler. Make sure that the vial
positions entered in this column are identical
to the vial positions used at the autosampler.
Inj-s
Number of injections for the same vial position
(only 1 injection is permissible for
Level ≥≥≥≥ 1
).
This value will not be transferred automatically
to the autosampler. Make sure that the
numbers of injections entered in this column
are identical to the numbers of injections
used at the autosampler.
Started
Indication whether the sample has been started
or not:
0
sample not started
1
sample started
2…
…
…
…n
number of injections done