beautypg.com

5 calibration and quantification – Metrohm 790 Personal IC User Manual

Page 109

background image

4 Operation

790 Personal IC

100

4.4.5

Calibration and quantification

General information

The aim of any chromatographic analysis is to answer the question

"What components are present in the sample and what are their con-

centrations?". Two procedures are used to achieve this goal: the first

step is called

calibration, the second step includes quantification.

Calibration has two aims: to get retention characteristics for all com-

ponents of interest (these data are stored in the component table) and

to establish a relation between injected amounts and corresponding in-

strumental responses for all components of interest (stored in the con-

centration table). Calibration is performed by running one or several

chromatograms of samples with known composition and known con-

centration of components (standards). For each calibrated component

a

calibration curve is constructed as a result.

With the 790 Personal IC three different procedures can be used for the

construction of the calibration curve. By far the most important method

for ion chromatography is the

external standard calibration (abso-

lute calibration) that is described in detail in this section. The other

methods of

internal standard calibration (relative calibration) and

tabulated calibration (relative gradient factor, a modified method for

external standard calibration) are of lesser importance and are not de-

scribed in detail here (for details please refer to on-line help).

Identification is a procedure that enables to decide what peaks on the

chromatogram correspond to what components. The identification is

performed based on the

Component table

created for calibration.

Quantification is a calculation procedure that determines components

concentrations, based on instrumental response (peak height or area),

using the calibration curves obtained earlier for each component.