3 connection of supply bottles – Metrohm 790 Personal IC User Manual

Page 49

background image

2 Installation

790 Personal IC

40

40

40

40

40 Tubing cartridge (6.2755.000)

for pump tubings 74/75

74/75

74/75

74/75

78

78

78

78 Suppressor inlet capillary for

eluent

41

41

41

41 Contact pressure lever

for adjusting the contact pressure

79

79

79

79 Suppressor outlet capillary for

eluent

42

42

42

42 Holding clamp

for locking the tubing cartridge into

place

80

80

80

80 Suppressor inlet capillary for H

2

O

43

43

43

43 Snap-action lever

for releasing the tubing cartridge

81

81

81

81 Suppressor inlet capillary for

H

2

SO

4

44

44

44

44 Pump drive

Roller head with contact rollers

82

82

82

82 Suppressor outlet capillary for H

2

O

45

45

45

45 Mounting pin

for attaching the tubing cartridge

83

83

83

83 Suppressor outlet capillary for

H

2

SO

4

2.8.3

Connection of supply bottles

The supply lines for the regeneration and rinsing solution between the

storage bottles and the peristaltic pump are installed as follows:

1 Prepare supply bottle for H

2

SO

4

• Prepare regeneration solution suited for the desired applica-

tion and separating column (normally 20 mmol/L H

2

SO

4

).

• Fill regeneration solution into supply bottle and label the

bottle.

2 Connect aspirating tubing for H

2

SO

4

• Prepare aspirating tubing 72

72

72

72: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally ca. 120 cm).

• Insert one end of the aspirating tubing 72

72

72

72 into the regenera-

tion solution storage bottle and tighten it so that the tubing is

firmly held.

• Insert the free end of aspirating tubing 72

72

72

72 into one of the

openings 5

555 of the 790 Personal IC (see Fig. 2) and pull it suf-

ficiently far into the interior.

• Mount a coupling 73

73

73

73 (6.2744.030) to the inlet end of the rear

pump tubing 74

74

74

74.

• Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 72

72

72

72 and screw this compression fitting on to

the coupling 73

73

73

73 (see Fig. 15).

3 Prepare supply bottle for H

2

O

• Prepare rinsing solution suited for the desired application and

separating column (normally dist. H

2

O).

• Fill rinsing solution into a supply bottle and label the bottle.