beautypg.com

4 connection of peek capillaries – Metrohm 790 Personal IC User Manual

Page 26

background image

2.3 Attaching the accessories

790 Personal IC

17

2.3.4

Connection of PEEK capillaries

For the connections between high-pressure pump and detector block

6.1831.010 PEEK capillaries (i.d. = 0.25 mm, e.d. =

1

/

16

"

) are used

which are connected using either

6.2744.010 PEEK compression fit-

tings (long) or 6.2744.070 PEEK compression fittings (short).

These PEEK connectors can also be used to connect 6.1822.010 PTFE

microcapillaries (i.d. = 0.3 mm). Proceed as follows:

Capillaries fitted with new connectors must have a perfectly flat cut

surface. To cut PEEK or PTFE capillaries it is best to use the

6.2621.080 Capillary tubing cutter.

1 Mount compression fitting

Slide a compression fitting 46

46

46

46 (6.2744.010) or a compression

fitting 47

47

47

47 (6.2744.070) over the end of the capillary 48

48

48

48 to be

fastened as shown in Fig. 7.

2 Insert capillary in connection

Push capillary end in the corresponding connection as far as it

will go (to avoid dead volume).

3 Tighten compression fitting

Tighten compression fitting 46

46

46

46 or 47

47

47

47 by hand (never use tools).



46

46

46

46 Compression

fitting

(6.2744.010)

47

47

47

47 Compression fitting

(6.2744.070)

48

48

48

48 Capillary

6.1831.010 PEEK capillary or

6.1822.010 PTFE microcapillary

46

46

46

46

48

48

48

48

47

47

47

47

Fig. 7:

Connectors for capillaries