3.3 Calibration
790 Personal IC
55
9 Switch injection valve to "FILL" position
⇒ Click on the
button of the 790 icon in the system
window to switch the injection valve to the "FILL" position. At
the same time, the suppressor module is switched to the next
position.
10 Fill sample loop
⇒ Immerse the aspirating tubing 23
23
23
23 in the standard solution.
⇒ Using the syringe fixed to syringe tubing 25
25
25
25, siphon in ca.
1 mL standard solution.
11 Switch injection valve to "INJECT" position
⇒ Click on the
button of the 790 icon in the system
window to switch the injection valve to the "INJECT" position.
At the same time, the data recording is started automatically.
The status bar of the chromatogram window shows the mes-
sage
Measure
, the
SYSTEM STATE
window shows the mes-
sage
INJECT done [790COM#]
.
⇒ When the analysis time of 14 min set in the method has
expired, the recorded chromatogram is automatically evalu-
ated and stored. The chromatogram window then closes.
12 Open chromatogram
⇒ Click on
or
File / Open / Chromatogram
in the main win-
dow. Select the chromatogram file
*.chw
just recorded and
click on
. The chromatogram window is opened where
the found peaks are numbered and the baselines are drawn
in.