Metrohm 790 Personal IC User Manual
Page 104
4.4 Methods
790 Personal IC
95
European Pharmacopoeia
Formula setting according to European Phar-
macopoeia.
US Pharmacopoeia
Formula setting according to US Pharmaco-
poeia.
Void time/volume
Calculation of
void time/void volume:
Calculation method
Selection of the method for void time calcula-
tion:
None
Manual entry in the
Void time
field.
1st
component
The peak corresponding to the first component
is selected as a void time marker and its reten-
tion time replaces the previous value of the void
time. If the first component is not identified, the
expected retention time is used.
1st peak
The first detected peak is used as a void time
marker for the run and its retention time is
stored in the
Void time
field.
From void volume %
For the calculation of the void time, the % value
entered in the
Void volume
field is multiplied by
the ratio of empty column volume and eluent
flow rate.
Void volume
Void volume in mL and % of column volume.
The % value can be entered manually if
Calcula-
tion method =
From void volume %
is set.
Void time
Void time in s. This value can be entered
manually if
Calculation method =
None
is set.
Index
Calculation of
Retention index:
Interpolation
Use of linear or logarithmic interpolation scale
for retention index calculation.
Internal
The index scale is constructed on the basis of
the current chromatogram. All components
used for index scale calibration should be pre-
sent in the current sample.
External
The index scale is constructed based on an-
other standard chromatogram.