2 Installation
790 Personal IC
30
2.7
Precolumns and separating columns
2.7.1
General information on precolumns
The use of easily exchangeable precolumns protects the separating
columns and prolongs their lifetime. The precolumns available from
Metrohm (see section 6.3.2) are either real precolumns or precolumn
cartridges, which are used together with the 6.2821.040 Cartridge head
or the 6.2828.010 Precolumn cartridge holder.
New IC precolumns are normally filled with solution and sealed at
both ends. Before the precolumn is installed in the system, it must be
ensured that this solution is freely miscible with the eluent used
(check manufacturer's specifications).
2.7.2
Precolumns with cartridge head
The 6.1005.020, 6.1005.040, 6.1005.050, 6.1007.010 and 6.1010.010
Precolumn cartridges are installed in the 6.2821.040 Cartridge head as
follows (see Fig. 11):
1 Prepare separating column
• Remove end caps from separating column 63
63
63
63.
• Unscrew fastening screw from column inlet.
• Take steel connector 56
56
56
56 for ferrule out of fastening screw.
2 Install cartridge
• Remove end caps for precolumn cartridge 59
59
59
59 (the steel mesh
58
58
58
58 and gaskets 57
57
57
57 are already installed in the cartridge).
• Mount steel spacer 60
60
60
60 on Manufit pressure screw 62
62
62
62 (the
steel mesh 61
61
61
61 and gaskets 57
57
57
57 are already installed in the
pressure screw).
• Mount precolumn cartridge 59
59
59
59 on the steel spacer 60
60
60
60 (com-
ply with flow direction if specified on the precolumn).
• Screw Manufit pressure screw 62
62
62
62 firmly to separating column
63
63
63
63.
• Mount steel connector 56
56
56
56 for ferrule on the inlet side of the
precolumn cartridge 59
59
59
59.
• Screw on Manufit housing 55
55
55
55 with Manufit pressure screw 62
62
62
62.