Metrohm 790 Personal IC User Manual
Page 115
4 Operation
790 Personal IC
106
Concentration table
790 PERSONAL IC / Method / Calibration / Concentrations
Selecting this menu item or clicking on
in the compo-
nent table opens the
Concentrations
window with the concentration table
containing concentrations of all components for all
Calibration levels
.
Each calibration level corresponds to a sample used for calibration and
to a point on the calibration curve.
Concentration units
User defined units of concentration that will ap-
pear in the report. Nevertheless, entry of new
concentration units does not cause recalcula-
tion of concentrations.
Data type
Choice of the data type that will be shown in the
concentration table:
concentrations
(default set-
ting), peak
heights
, peak
areas
,
volumes
.
Number
Number of the component (read-only).
Name
Name of the component taken from the com-
ponent table (read-only).
This run
Contains concentrations (or other chosen val-
ues) obtained in the current run (read-only, ex-
cept for universal component).
Level 1...Level N
Contains concentrations (or other chosen val-
ues) of components for corresponding
Calibra-
tion level
.