5.2 Maintenance and servicing

790 Personal IC

167

5.2.10 Exchanging the pump tubing

The pump tubings used by the peristaltic pump are consumable mate-

rial with a limited lifetime and should be exchanged at regular intervals

(approx. every 4 weeks under continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in section 2.9.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges 40

40

40

40 should

be raised completely by loosening the snap-action lever 43

43

43

43 on the

right-hand side (the set contact pressure remains unchanged).
As the pump is always operated on the same side, the 6.1826.060

Pump tubings supplied can be used on both sides. To exchange a

pump tubing proceed as follows:

1 Remove old pump tubing

• Press contact pressure lever 41

41

41

41 on the tubing cartridge 40

40

40

40

down as far as it will go.

• Release tubing cartridge 40

40

40

40 from holding clamp 42

42

42

42 by press-

ing down snap-action lever 43

43

43

43 and remove from mounting

pin 45

45

45

45 (see Fig. 15).

• Remove old pump tubing 74

74

74

74 or 75

75

75

75.

2 Insert new pump tubing

• Insert the new pump tubing 74

74

74

74 or 75

75

75

75 in the tubing cartridge

40

40

40

40 as shown in Fig. 14. The white-yellow stopper 76

76

76

76 must

click into the corresponding holder on the left-hand side of

the tubing cartridge.

• Place the tubing cartridge 40

40

40

40 on mounting pin 45

45

45

45 and press

down on the right-hand side until snap-action lever 43

43

43

43 clicks

into position on holding clamp 42

42

42

42. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

• Switch on peristaltic pump.
• Press contact pressure lever 41

41

41

41 upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.

• Switch off peristaltic pump.