2 Installation

790 Personal IC

18

2.3.5

Filter unit PEEK

One

6.2821.100 Filter unit PEEK (see Fig. 8) is already installed be-

tween the high-pressure pump and the injection valve 22

22

22

22 at the 790

Personal IC. This filter unit serves to avoid contamination by abrasive

particles of the piston seals.
The two other filter units PEEK supplied with the 2.790.0020 Personal IC

(with suppressor) are installed between the pump tubings of the peri-

staltic pump and the inlet capillaries for regeneration and rinsing solu-

tion (see section 2.8.2). These filter units serve to protect the suppres-

sor module from foreign particles and bacterial growth.
The Filter unit PEEK 28

28

28

28 consists of the housing 50

50

50

50 and the two connec-

tors 49

49

49

49 (with filter) and 51

51

51

51 (without filter) screwed into the housing 50

50

50

50.

For the connection of capillaries 48

48

48

48 PEEK compression fittings 46

46

46

46

(6.2744.010) or 47

47

47

47 (6.2744.070) must be used. New connectors 49

49

49

49 with

filter are available as an option with the ordering number 6.2821.110

(set of 10).

For the connection of the filter unit, please note the flow direction

arrow printed on the housing.

48

48

48

48

46

46

46

46

49

49

49

49

50

50

50

50

51

51

51

51

46

46

46

46

48

48

48

48

28

28

28

28

Fig. 8: 6.2821.100 Filter unit PEEK

28

28

28

28 Filter unit PEEK (6.2821.100)

49

49

49

49 Connector with filter (6.2821.110)

Part of 6.2821.100 Filter unit

46

46

46

46 Compression fitting (6.2744.010) 50

50

50

50 Housing for filter unit

Part of 6.2821.100 Filter unit

48

48

48

48 Capillary

6.1831.010 PEEK capillary or

6.1822.010 PTFE microcapillary

51

51

51

51 Connector without filter

Part of 6.2821.100 Filter unit