1 general description – Bowers Group Sylvac D100S Readouts User Manual
Page 71

70
5 PNEUMATIC COMMAND UNITS D110 AND D111 or VACUUM D110V and D111V
5.1 GENERAL DESCRIPTION
The pneumatic command units D110/D111 or D110V/D111V (vacuum) are designed to work with the display units
D100S. They can be mounted on the display unit by inserting the 4 plastic legs, or, for weight considerations, under the
display unit. They can also be positioned separately away from the display unit, close to the probes.
- the D110/D110V units allow the control of pneumatic/vacuum lifting of from 1 to 16 probes.
- the D111/D111V pneumatic/vacuum extension units allows pneumatic/vacuum lifting of 16 additional probes.
A maximum of 1 D110/V and 3 D111/V units can be connected to one D100S, allowing the pneumatic/vacuum com-
mand of a maximum of 64 probes.
The retraction and return motion of probes is commanded from the D100S unit, i.e. by the
and
,
keys or by means of the foot pedal. Remote command from a computer for pneumatic
lifting is equally possible. Please note that all pneumatic lifters or vacuum probes are commanded simultaneously, so
it is not possible to control each lifter separately.
5.2 REAR PANEL (D110 unit)
5.3 OPERATION OF D110 UNIT
1/ Connect the D110 unit to the D100S unit ("Output" socket) by means of the 15 pin connecting cable.
2/ Connect from 1 to 16 lifters to the pneumatic outputs of the D110 unit. Use the semi-flexible black PUR tube with an
outside diameter of 4mm and an inside diameter of 2.5mm which is supplied with the pneumatic lifter. Unused soc-
kets must have the red plastic plug which close the air outlet.
3/ Connect the air system at the point marked 'In' on the unit using the quick connector supplied and a tube with an out-
side diameter of 6mm and an inside diameter of 4mm. Filtered and dry air, 3 to 6 bar.
4/ Configure the D100S unit for lifting:
- by means of the
and
keys: Set Up then Enter (refer to Section 1.6.2).
- using the foot pedal (external contact):
then
then 1 then 6 (refer to Section 1.6.11).
- using the computer: this gives the UP order for lifting and DOW (DOWN) for the return motion (refer to Section 1.6.14:
remote command).
5/ Control the probe return speed by means of the microflow restrictor thumbscrew on the rear panel (can be locked by
means of the locknut).
From Outputs D100
Sylvac SA CH-1023 Crissier Swiss made
Ser.Nr
Out
In
3 to 6 Bars
Up to 16 Lifters