1.11 REPLACING ACCUMULATORS

When the autonomy of the D100S unit is no longer satisfactory it is necessary to replace the accumulators inside the
unit (their lifespan is approx. 4 years).

Replacement of accumulators can be carried out either by a Sylvac representative or as follows:

- Obtain a new battery pack for the D100S from your Sylvac agent.

- Open the unit as described in Section 1.10.

- Remove the old accumulator connector by pulling it back:

- Loosen two fixing screws on the battery pack bracket.

- Install the new accumulator as shown in the illustration:

- Replace the two fixing screws on the bracket. Plug in the battery pack connector.

- Replace the lower half of the housing, ensuring that the front and back panels are correctly positioned in the specially

designed grooves.

- Replace the 4 cover screws.

- Plug the mains charger into the unit and charge the accumulators for at least 15 hours.