Bowers Group Sylvac D100S Readouts User Manual
Page 22
![background image](/manuals/584793/22/background.png)
1.6.8 CLASS KEY
Displays or removes the classification of measured values.
Indicates measured value under the 1st class limit.
Indicates measured value inside class 3 (n = 2 to 6).
Indicates measured value over the last class limit.
This information is also sent to the corresponding optocoupler for external command purposes (refer to Section 1.8.3)
- Input of classes:
1/ press
2/ then
The following will be displayed:
3/ enter the required number of classes, minimum 2 and maximum 6 (1 class
is equivalent to tolerance mode).
4/ then enter the lower limit of class 1, e.g. 12.4 mm. Numerical values are
entered as described in Section 1.3.3.
5/ enter the upper limit of class 1, e.g. 12.6 mm.
6/ then enter the upper limit of class 2, e.g. 12.8 mm.
7/ and so on, up to the selected number of classes.
Once this has been done, the unit reverts back to measuring mode. The measured value and the class graphics will be
displayed. The measured value is displayed in reverse video on the corresponding class symbol.
In the above example, the limit of the 1st class is from 12.4 to 12.6 mm, the 2nd class is from 12.6 to 12.8 mm, etc.
Measured values under the lower limit, e.g. 12.4 will be shown by the symbol
, measured values over the upper
limit, e.g. 12.8, by the symbol
.
By pressing
the graphic symbols will be cancelled.
Pressing
again displays them again, with the parameters
and
previously entered.
>
<
CHANNEL 1
Enter number of
classes (2 à 6):
>
3
<
21